The use of dual accelerometers improves measurement of chest compression depth
Jaehoon Oh1, Yeongtak Song, Boseung Kang
1Dept. of Emergency Medicine, College of Medicine, Hanyang University, Seoul, Republic of Korea.
Background:
Chest compression (CC) feedback devices are used to perform CC measurements effectively and accurately on patients in hospital beds. However, these devices do not take account of the compression of the mattress, which results in overestimation of CC depth. In this study, we propose a new method using two accelerometers to overcome this limitation and thus measure compression depth more accurately when performing cardiopulmonary resuscitation (CPR) on patients.
Method:
One accelerometer was placed on the manikin's sternum (a1), and the other between the manikin's back and the mattress (a2). The compression depth was calculated by integrating the acceleration twice using a digital signal processing technique. We compared CC depth from dual accelerometers and single accelerometer (a1) on the foam and inflated air mattress with eight CPR providers.
Result:
When CC was done on a manikin lying on the floor, there was no significant difference between measurement techniques (p>0.05). When CC was done on a manikin lying on the foam and inflated air mattress supporting system, our method significantly improved the estimation of CC depth, irrespective of the presence or absence of a backboard (p<0.001).
Conclusion:
Measuring CC depth using two accelerometers is more effective than using one in increasing the accuracy of CC depth estimation when CPR is performed on the foam and inflated air mattress, regardless of the presence or absence of a backboard.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Measurement of Blood Pressure
Special considerations while measuring pulse
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...