Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
An evaluation of peak inspiratory pressure, tidal volume, and ventilatory frequency during ventilation with a
Mariana Almada Bassani1, Francisco Mezzacappa Filho, Maria Regina de Carvalho Coppo
1Universidade Estadual de Campinas, Campinas, São Paulo, Brazil. bassanimariana@hotmail.com
Background:
Although the self-inflating bag is widely used in the hospital setting, variability of delivered ventilatory parameters is usually high, which might result in both hypoventilation and lung injury. The aims of this study were to assess possible sources of the high variability and to evaluate the adequacy of obtained values in relation to the recommended values for neonatal resuscitation.
Methods:
This was an experimental study in which 172 health professionals (physicians, resident physicians, physiotherapists, nurses, and nursing technicians) who work with neonatal intensive care manually ventilated a test lung (adjusted to simulate the lungs of an intubated term newborn) with a self-inflating bag in 5 different handling techniques, using 10, 5, 4, 3, and 2 fingers. Delivered values of peak inspiratory pressure (PIP), tidal volume (V(T)), and ventilatory frequency (f) were compared, taking into account the different handling modalities and professions by analysis of variance for repeated measures. Chi-square, the Friedman test and the Fisher exact tests were performed to compare the delivered and standard values.
Results:
PIP and V(T) were significantly affected by the handling technique, with higher values for a greater number of fingers used for ventilation. Profession also influenced V(T) and f significantly: physiotherapists tended to deliver higher volumes and lower rates. Nevertheless, we observed high variability of all studied ventilatory parameters and overall inadequacy of obtained values. Most volunteers delivered excessive pressures and volumes at insufficient ventilatory frequency.
Conclusions:
Delivered values seem to depend on operators' individual and professional differences, as well as on the number of fingers used to compress the bag. However, from the clinical point of view, it is important to point out the high occurrence of inadequate delivered values, regardless of handling technique and profession.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Respiratory Volumes and Capacities
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)