Use of a single ventilator to support 4 patients: laboratory evaluation of a limited concept
Richard D Branson1, Thomas C Blakeman, Bryce Rh Robinson
1Department of Surgery, University of Cincinnati, Cincinnati, OH, USA. richard.branson@uc.edu
Introduction:
A mass-casualty respiratory failure event where patients exceed available ventilators has spurred several proposed solutions. One proposal is use of a single ventilator to support 4 patients.
Methods:
A ventilator was modified to allow attachment of 4 circuits. Each circuit was connected to one chamber of 2 dual-chambered, test lungs. The ventilator was set at a tidal volume (V(T)) of 2.0 L, respiratory frequency of 10 breaths/min, and PEEP of 5 cm H(2)O. Tests were repeated with pressure targeted breaths at 15 cm H(2)O. Airway pressure, volume, and flow were measured at each chamber. The test lungs were set to simulate 4 patients using combinations of resistance (R) and compliance (C). These included equivalent C and R, constant R and variable C, constant C and variable R, and variable C and variable R.
Results:
When R and C were equivalent the V(T) distributed to each chamber of the test lung was similar during both volume (range 428-442 mL) and pressure (range 528-544 mL) breaths. Changing C while R was constant resulted in large variations in delivered V(T) (volume range 257-621 mL, pressure range 320-762 mL). Changing R while C was constant resulted in a smaller variation in V(T) (volume range 418-460 mL, pressure range 502-554 mL) compared to only C changes. When R and C were both varied, the range of delivered V(T) in both volume (336-517 mL) and pressure (417-676 mL) breaths was greater, compared to only R changes.
Conclusions:
Using a single ventilator to support 4 patients is an attractive concept; however, the V(T) cannot be controlled for each subject and V(T) disparity is proportional to the variability in compliance. Along with other practical limitations, these findings cannot support the use of this concept for mass-casualty respiratory failure.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Pneumonia IV: Management
Bacterial Pneumonia Treatment
For bacterial pneumonia, antibiotics serve as the cornerstone of therapy. Initial treatment often begins with empirical antibiotics, tailored to the anticipated causative organism and adjusted based on culture results. Key antibiotic choices include:
Acute Respiratory Failure-IV