Intravenous pyelogram artefacts unique to digital tomosynthesis reconstruction
1Clinical and Radiation Physics Department, Plymouth Hospitals NHS Trust, Derriford Road, Plymouth PL6 8DH, UK. ben.rowberry@phnt.swest.nhs.uk
Digital tomosynthesis (DT) intravenous pyelogram (IVP) can cause unique imaging artifacts due to patient breathing during acquisition. To prevent misdiagnosis, patients must perform adequate breath-holds for accurate DT IVP results.
Area of Science:
- Medical Imaging
- Radiology
- Diagnostic Technology
Background:
- Digital tomosynthesis (DT) offers retrospective multi-planar reconstruction for investigations like intravenous pyelogram (IVP).
- DT imaging utilizes a limited angular range, resulting in incomplete data sets and potential out-of-focus information in reconstructed images.
Purpose of the Study:
- To describe and identify the causes of two artifacts unique to DT imaging.
- To investigate the potential for misdiagnosis of renal system anatomy due to these artifacts.
Main Methods:
- Analysis of DT imaging data from intravenous pyelogram (IVP) examinations.
- Correlation of imaging artifacts with patient breathing, renal anatomy, acquisition methods, and reconstruction algorithms.
Main Results:
- Two novel artifacts unique to DT imaging were identified, not solely attributable to blurring.
- These artifacts are hypothesized to result from patient breathing during data acquisition, compounded by other factors.
- A potential for misdiagnosis of a duplex collecting system by unaware radiologists was noted.
Conclusions:
- DT IVP artifacts are likely caused by breathing during data acquisition combined with specific imaging parameters.
- Recommendations include using DT IVP only in patients capable of breath-holding.
- Performing DT IVP with a breath-hold after expiration is suggested to mitigate artifacts.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
05:41Synchronous Triplanar Reconstruction Integrated with Color Doppler Mapping for Precise and Rapid Localization of Thyroid Lesions
Published on: February 9, 2024
Related Concept Videos
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies VII: Vascular Imaging
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Imaging Studies VI: Voiding Cystourethrography and Cystography