Related Experiment Video
Updated: May 28, 2026

Precise Electrochemical Sizing of Individual Electro-Inactive Particles
Published on: August 4, 2023
Achieving nanometer scale tip-to-substrate gaps with micrometer-size ultramicroelectrodes in scanning electrochemical
Mei Shen1, Netzahualcóyotl Arroyo-Currás, Allen J Bard
1Center for Electrochemistry, Department of Chemistry and Biochemistry, University of Texas at Austin, 1 University Station, A5300, Austin, Texas 78712-0165, USA.
Abstract:
Scanning electrochemical microscopy (SECM) tips with rounded glass insulation around the metal wire (radius a = 5 μm) were fabricated (apparent RG < 1.1, where RG is the ratio of the radius of the insulation sheath divided by the electrode radius), and their SECM feedback approach curves were studied in solutions of tris(2,2'-bipyridine)ruthenium(2+) (Rubpy) in acetonitrile and ferrocenemethanol in water with a platinum disk as the substrate electrode (radius a(s) = 1 mm). Considerable enhancement of the normalized feedback current, I(T)(L) = i(T)/i(T,∞), where L = d/a and d is the distance traveled by the SECM tip, was observed in both systems (e.g., I(T)(L) = 15 in organic solutions and I(T)(L) = 30 in aqueous solutions) with good electrode alignment. This shows that tip-to-substrate gaps of ca. d = 110 nm can be achieved. To account for any deviations from the usual disk UME behavior and currents caused by possible changes in the tip electrode geometry, simulations of the feedback response were performed for a 2D axisymmetric environment. All simulated results match in a point-to-point comparison with experimental values (average relative standard deviation (RSD) = 0.01 ± 0.005).