Effective use of radiation shields to minimize operator dose during invasive cardiology procedures
Kenneth A Fetterly1, Dixon J Magnuson, Gordon M Tannahill
1Division of Cardiovascular Diseases, Mayo Clinic, Rochester, Minnesota 55905, USA. fetterly.kenneth@mayo.edu
Objectives:
This study sought to measure the protection from scatter radiation offered to the primary physician by a variety of available shields and to provide best practice guidelines for shield use during invasive cardiology procedures.
Background:
It is accepted that exposure to radiation includes a predicted increase in cancer risk. In the cardiac interventional laboratories, radiation shields are widely available; however, proper use of the shields to optimize protection during cardiac interventional procedures is not well understood.
Methods:
The protection from scatter radiation offered by a variety of shields used alone and in combination was measured. Protection was assessed from air-kerma measurements of scatter radiation from a phantom performed without and with the shields. Protection was assessed for 3 patient- access locations (right jugular vein, right femoral artery, and left anterior chest) and for elevations ranging from 25 to 175 cm from the floor. The influence of precise placement of the ceiling-mounted upper body shield was specifically assessed.
Results:
The utility and protection of shielding varied for the 3 access points and with elevation. For femoral artery access locations, the shields can provide at least 80% protection from scatter at all elevations; however, protection depends substantially on upper body shield position. A disposable radiation-absorbing pad can provide 35% to 70% upper body protection for procedures during which the upper body shield cannot be used effectively.
Conclusions:
Radiation shields can provide substantial protection from radiation during cardiac interventional procedures. Shields must be thoughtfully and actively managed to provide optimum protection. Best practice guidelines for shield use are provided.
More Related Videos
08:30X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization IV: Nursing Management
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...