Related Experiment Video
Updated: May 28, 2026

Expedited Radiation Biodosimetry by Automated Dicentric Chromosome Identification (ADCI) and Dose Estimation
Published on: September 4, 2017
Automated framework for digital radiation dose index reporting from CT dose reports
George Shih1, Zheng Feng Lu, Ramin Zabih
1Department of Radiology, Weill Cornell Medical Center, New York, NY 10065, USA. ges9006@med.cornell.edu
A new Dose Index Reporting Application (DIRA) accurately extracts radiation dose index information from CT scans. This system enables automated quality control and longitudinal tracking of patient radiation exposure, enhancing safety awareness.
Area of Science:
- Medical Imaging
- Radiology
- Health Informatics
Background:
- Computed tomography (CT) radiation exposure has risen significantly, comprising 50% of medical radiation.
- Current CT dose index (CTDI) tracking is manual and difficult due to non-digital, image-embedded data.
- Lack of automated systems hinders quality control and patient radiation exposure monitoring.
Purpose of the Study:
- To develop a Dose Index Reporting Application (DIRA) for CT.
- To automate the extraction of radiation dose index information from CTDI reports.
- To enable automated quality control, radiation safety awareness, and longitudinal patient radiation exposure records.
Main Methods:
- Developed a vendor-independent DIRA system for CT scanners.
- Processed 518 CTDI reports using DICOM C-STORE for automated data extraction.
- Utilized a web-based framework with a SQL database for compiling dose and patient information.
Main Results:
- DIRA achieved 100% accuracy in extracting dose index information from 518 CTDI reports.
- Developed correction factors for patient size using variable phantom sizes.
- Demonstrated automated CTDI tracking on existing CT scanners, independent of DICOM SR Dose Index Report standard.
Conclusions:
- The modular DIRA system integrates with any CT scanner, facilitating digital dose index reporting.
- Enables a longitudinal record of healthcare radiation exposure in individual patient records.
- Promotes enhanced radiation safety and automated quality control in CT imaging.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
05:18Radiation Planning Assistant - A Web-based Tool to Support High-quality Radiotherapy in Clinics with Limited Resources
Published on: October 6, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Biological Effects of Radiation