Patient-specific three-dimensional concomitant dose from cone beam computed tomography exposure in image-guided
Emiliano Spezi1, Patrick Downes, Richard Jarvis
1Department of Medical Physics, Velindre Cancer Centre, Cardiff, United Kingdom. Spezi@wales.nhs.uk
Purpose:
The purpose of the present study was to quantify the concomitant dose received by patients undergoing cone beam computed tomography (CBCT) scanning in different clinical scenarios as a part of image-guided radiotherapy (IGRT) procedures.
Methods And Materials:
We calculated the three-dimensional concomitant dose received as a result of CBCT scans in 6 patients representing different clinical scenarios: two pelvis, two head and neck, and two chest. We assessed the effect that a daily on-line IGRT strategy would have on the patient dose distribution, assuming 40 CBCT scans throughout the treatment course. The additional dose to the planning target volume margin region was also estimated.
Results:
In the pelvis, a single CBCT scan delivered a mean dose to the femoral heads of 2-6 cGy and the rectum of 1-2 cGy. An additional dose to the planning target volume was within 1-3 cGy. In the chest, the mean dose to the planning target volume varied from 2.5 to 5 cGy. The lung and spinal cord planning organ at risk volume received ≤4 cGy and ≤5 cGy, respectively. In the head and neck, a single CBCT scan delivered a mean dose of 0.3 cGy, with bony structures receiving 0.5-0.8 cGy. The femoral heads received an additional dose of 1.5-2.5 Gy. A reduction of 20-30% in the mean dose to the organs at risk was achieved using bowtie filtration. In the head and neck, the dose to the eyes and brainstem was eliminated by decreasing the craniocaudal field size.
Conclusions:
The additional dose from on-line IGRT procedures can be clinically relevant. The organ dose can be significantly reduced with the use of appropriate patient-specific settings. The concomitant dose from CBCT should be accounted for and the acquisition settings optimized for optimal IGRT strategies on a patient basis.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography