Injection force of reusable insulin pens: Novopen 4, Lilly Luxura, Berlipen, and ClikSTAR
Arnd Friedrichs1, Volker Korger, Steffen Adler
1LWS Risk Management Consult, Brannenburg, Germany. arnd.friedrichs@lwsgroup.com
Background:
Insulin pen devices are used by approximately half of insulin users worldwide. The injection force of insulin pens is a key element in their design. This study aimed to demonstrate that the sanofi-aventis reusable ClikSTAR® (CS) pen has an improved injection force over existing insulin pens.
Methods:
The injection force of four reusable insulin pens--Novopen® 4 (NP4; Novo Nordisk), Luxura® (LL; Eli Lilly and Co.), Berlipen® (BP; Haselmeier GmbH), and CS (sanofi-aventis)--was tested in a laboratory setting. Injection force was tested using two methods: six dispense rates between 6 and 24.66 U/s (constant volume flow rate) and constant button speeds of 4 and 8 mm/s.
Results:
The CS required a lower mean injection force versus NP4, LL, and BP at both doses and all dispense rates. Mean injection force was 45%, 126%, and 60% higher for NP4, LL, and BP versus CS, respectively (p < .05 for each of the comparisons), for a flow rate of 6 U/s at 60 U dose. Mean injection force in all pens increased with the dispense rate, but the injection force remained significantly lower for CS versus all other pens (p < .05). The injection force for CS was significantly lower for 60 U at 10 and 17.03 U/s than for 80 U.
Conclusions:
The study demonstrated that CS pens require a lower injection force at a wide range of different injection speeds than other reusable insulin pens. This is an important benefit for patients with diabetes, especially those with limited dexterity.
Related Concept Videos
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into rapid-acting...
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Production of Pharmaceuticals
Insulin: Biosynthesis, Chemistry, and Preparation
Damage or functional impairment of β-cells inhibits insulin production, leading to diabetes. Diabetes treatment primarily uses...
Insulin Secretory Vesicles