Related Experiment Video
Updated: May 28, 2026

Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
Respiratory rate extraction from pulse oximeter and electrocardiographic recordings
Jinseok Lee1, John P Florian, Ki H Chon
1Department of Biomedical Engineering, Worcester Polytechnic Institute, Worcester, MA, USA. jinseok@wpi.edu
Abstract:
We present an algorithm of respiratory rate extraction using particle filter (PF), which is applicable to both photoplethysmogram (PPG) and electrocardiogram (ECG) signals. For the respiratory rate estimation, 1 min data are analyzed with combination of a PF method and an autoregressive model where among the resultant coefficients, the corresponding pole angle with the highest magnitude is searched since this reflects the closest approximation of the true breathing rate. The PPG data were collected from 15 subjects with the metronome breathing rate ranging from 24 to 36 breaths per minute in the supine and upright positions. The ECG data were collected from 11 subjects with spontaneous breathing ranging from 36 to 60 breaths per minute during treadmill exercises. Our method was able to accurately extract respiratory rates for both metronome and spontaneous breathing even during strenuous exercises. More importantly, despite slow increases in breathing rates concomitant with greater exercise vigor with time, our method was able to accurately track these progressive increases in respiratory rates. We quantified the accuracy of our method by using the mean, standard deviation and interquartile range of the error rates which all reflected high accuracy in estimating the true breathing rates. We are not aware of any other algorithms that are able to provide accurate respiratory rates directly from either ECG signals or PPG signals with spontaneous breathing during strenuous exercises. Our method is near real-time realizable because the computational time on 1 min data segment takes only 10 ms on a 2.66 GHz Intel Core2 microprocessor; the data are subsequently shifted every 10 s to obtain near-continuous breathing rates. This is an attractive feature since most other techniques require offline data analyses to estimate breathing rates.
Related Concept Videos
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Respiratory Volumes and Capacities I