Detection and Removal of Tooth-Colored Composite Resin Using the Fluorescence-Aided Identification Technique
Published on: July 27, 2022
Light-emitting diode and laser fluorescence-based devices in detecting occlusal caries
Jonas A Rodrigues1, Isabel Hug, Klaus W Neuhaus
1Federal University of Rio Grande do Sul (UFRGS), School of Dentistry, Department of Pediatric Dentistry, Rua Ramiro Barcelos 2492, 90035-003, Porto Alegre, Brazil. jorodrigues@ufrgs.br
This study evaluated laser fluorescence (LF) and light-emitting diode (LED) devices for detecting tooth decay. LF devices showed promise as auxiliary tools, while one LED device struggled with early enamel caries detection.
Area of Science:
- Dental diagnostics
- Caries detection technology
- In vitro studies
Background:
- Accurate detection of occlusal caries is crucial for timely intervention.
- Conventional methods have limitations in early caries detection.
- Novel diagnostic devices, including LED and laser fluorescence, are being developed.
Purpose of the Study:
- To compare the diagnostic performance of two LED devices (Midwest Caries - MID, VistaProof - VP) and two laser fluorescence (LF) devices (DIAGNOdent 2095, DIAGNOdent pen 2190) for detecting occlusal caries in vitro.
- To evaluate the accuracy, specificity, and sensitivity of these devices against histological findings.
- To assess inter- and intra-examiner reliability.
Main Methods:
- Ninety-seven permanent molars were examined twice by two examiners.
- Devices used: two LED (MID, VP) and two LF (LF, LFpen).
- Teeth were histologically prepared and classified for lesion depth (D1 and D3 thresholds).
- Statistical analysis included Spearman's rank correlation and ICC for reliability.
Main Results:
- At the D3 threshold, LF devices showed specificities of 0.88 (LF) and 0.87 (LFpen), and sensitivities of 0.63 for both.
- The MID device demonstrated high specificity (0.94 at D1, 0.90 at D3) but lower sensitivity (0.31 at D1, 0.70 at D3).
- The VP device showed higher sensitivity (0.75 at D1, 0.96 at D3) but lower specificity (0.70 at both thresholds).
- Spearman correlations ranged from 0.51 to 0.58, indicating moderate agreement with histology.
- Inter- and intra-examiner reliability (ICC) was high (0.83-0.90).
Conclusions:
- Laser fluorescence devices are valuable auxiliary tools for caries detection, offering good reproducibility and improved accuracy at the D3 threshold.
- The MID device was insufficient for differentiating sound surfaces from early enamel caries.
- The VP device requires refinement of its cut-off limits for optimal clinical application.