Related Experiment Video
Updated: May 28, 2026

Evaluation of Right Ventricular Function in Experimental Models of Pulmonary Arterial Hypertension
Published on: June 27, 2025
Pulse transit time: validation of blood pressure measurement under positive airway pressure ventilation
Heidi Schmalgemeier1, Thomas Bitter, Stephan Bartsch
1Department of Cardiology, Heart and Diabetes Centre North Rhine-Westphalia, University Hospital, Ruhr-University Bochum, Georgstrasse 11, 32545, Bad Oeynhausen, Germany.
Purpose:
The objective of the present study was to validate noninvasive, continuous blood pressure (BP) measurement using pulse transit time (PTT) to represent absolute values and detect BP changes under continuous positive airway pressure (CPAP) treatment.
Methods:
We applied CPAP to 78 patients of a cardiological sleep lab using 0 (baseline), 4, 8, and 12 cmH(2)O for 10 min at every level and measured BP simultaneous via PTT-based method and standard oscillometric method (OM).
Results:
Quality of signal perception was acceptable to convert PTT into BP values in 64 patients (82%). When comparing both methods, we found a strong linear correlation of systolic and diastolic BP (baseline, r = 0.94 for systolic BP; r = 0.95 for diastolic BP, p < 0.001) while no significant differences between absolute values obtained with OM and PTT measurement. Mean bias at baseline was 4.1 ± 3.2 mmHg for systolic BP and 2.3 ± 2.2 mmHg for diastolic BP. With increasing CPAP levels, PTT and OM measurements differed continuously up to a systolic difference of 6.6 ± 4.9 mmHg and a diastolic difference of 4.4 ± 3.5 mmHg. There was no definite trend of PTT method to either over- or underestimate BP.
Conclusions:
We found that PTT- and OM-based BP results are closely correlated while applying CPAP over a period of 40 min. With higher CPAP level, bias between both methods increased.
Related Concept Videos
Measurement of Blood Pressure
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Pulse
The pulse serves as a clinical indicator...