Gamma-irradiation sterilization in an inert environment: a partial solution
Daniel J Berry1, Barbara H Currier, Michael B Mayor
1Department of Orthopedic Surgery, Mayo Clinic, Rochester, MN, USA.
Background:
In the mid to late 1990 s, to sterilize UHMWPE bearings, manufacturers changed from gamma-irradiation-in-air (gamma-air) sterilization, which initiated oxidation leading to bearing fatigue, to gamma-irradiation sterilization in an inert environment (gamma-inert). The change to gamma-inert sterilization reportedly prevented shelf oxidation before implantation but not in vivo oxidation.
Questions/Purposes:
We asked: (1) Has the change to gamma-inert sterilization prevented shelf oxidation that led to early in vivo fatigue damage in gamma-air-sterilized tibial inserts? And (2) has the change to gamma-inert sterilization prevented the occurrence of fatigue secondary to in vivo oxidation?
Methods:
We rated 183 retrieved gamma-air- and 175 retrieved gamma-inert-sterilized tibial inserts for clinical fatigue damage and analyzed 132 gamma-air- and 174 gamma-inert-sterilized tibial inserts for oxidation by Fourier transform infrared spectroscopy.
Results:
Oxidation led to decreased mechanical properties in shelf-aged gamma-air-sterilized tibial inserts. Barrier packaging prevented shelf oxidation in gamma-inert-sterilized tibial inserts. Gamma-air- and gamma-inert-sterilized inserts oxidized in vivo. Fatigue damage (delamination) occurred more frequently in inserts retrieved after longer time in vivo. Longer in vivo time correlated with higher oxidation and more accumulated cycles of use.
Conclusions:
Published oxidation projections suggest gamma-inert-sterilized tibial inserts would reach the critical oxidation for the onset of fatigue after 11 to 14 years in vivo. These retrievals appear to follow the projected oxidation trends. Frequency of fatigue damage increased with increasing oxidation.
Clinical Relevance:
Fatigue of tibial inserts becomes more likely, especially in active patients, after more than a decade of good clinical performance.
More Related Videos
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Temperature
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Key Techniques in Microbiology