Related Experiment Video
Updated: May 28, 2026

Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Organ dose evaluation for multi-slice spiral CT scans based on China Sichuan chest anthropomorphic phantom
Gang Peng1, Yongming Zeng, Tianyou Luo
1Department of Radiology, The First Affiliated Hospital of Chongqing Medical University, Chongqing 400016, China.
Abstract:
The authors measured organ radiation doses during multi-slice computed tomography (MSCT) chest scans using a China Sichuan anthropomorphic phantom (CDP-1C). Chest CT images from live volunteers based on automatic tube current modulation (ATCM) techniques were similar to those obtained using the CDP-1C phantom, indicating that the phantom accurately modelled the anatomic structure and X-ray absorbance of the human torso. Indeed, attenuation values differed by <5%. Organ radiation doses were measured using thermoluminescence dosemeters in the CDP-1C. With increased noise index, the CT dose index, the dose-length product and the average organ dose all decreased. Thus, the CDP-1C phantom can also assess dose levels during CT examinations in Chinese patients. The noise index (based on ATCM techniques) should be set to 8.5 or higher to reduce X-ray exposure while maintaining appropriate resolution for diagnosis.
