Radiation exposure in whole-body computed tomography of multiple trauma patients: bearing devices and patient positioning
Related Experiment Video
Updated: May 27, 2026

A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Björn Loewenhardt1, Michael Buhl, André Gries
1Center of Radiology, Department of Diagnostic and Interventional Radiology and Neuroradiology, Hospital of Fulda, Pacelliallee 4, D-3604 Fulda, Germany. bjoern.loewenhardt@web.de
Background:
Whole-body computed tomography (WBCT) plays an important role in the management of severely injured patients. We evaluated the radiation exposure of WBCT scans using different positioning boards and arm positions.
Methods:
In this retrospective study, the radiation exposure of WBCT using a 16-slice multislice computed tomography scanner was evaluated. Individual effective doses (E, mSV) was calculated. Patients were assigned to two groups according to placement on a plastic transfer mat (PTM, group 1) or on the Trauma Transfer™-Board (TTB, group 2). Data were collected for each group with arm placement on the abdomen (a) or in raising position (b), respectively. The maximum ventro-dorsal diameter [VDD] at the trunk was measured.
Results:
100 patients with potentially life-threatening injuries were analysed. Patient demographics and VDD did not differ in the two groups. Radiation exposure in term of E did not reveal any significant differences between the two positioning boards using same arm position [group 1a (n=26) vs. 2a (n=24) (mSV): 16.7±4.7 vs. 17.1±4.4, group 1b (n=26) vs. 2b (n=24) (mSV): 13.1±3.9 vs. 14.3±1.5]. The arm raising positioning showed a significant reduction in E in comparison to the placement on abdomen position [group 1b vs. 1a (mSV): 13.1±3.9 vs. 16.7±4.7, p<0.05, group 2b vs. 2a (mSV): 14.3±1.5 vs. 17.1±4.4, p<0.05].
Conclusions:
Patient arm positioning for WBCT has an important influence on radiation exposure. Effective dose was 16-22% lower when arms were raised. An individual placement algorithm may lead to a relevant reduction of radiation exposure of severely injured patients.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
X-ray Imaging
Biological Effects of Radiation
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET