Trueness in the measurement of haemoglobin: consensus or reference method?
Kristel J M Boonen1, Joyce Curvers, Alex A Timmerman
1Catharina Hospital Eindhoven, Clinical Laboratory, Eindhoven, The Netherlands. kristel.boonen@cze.nl
Haemoglobin measurements on Cell Dyn Sapphires show a negative bias. External quality assessment consensus values also deviate significantly from the haemiglobincyanide reference method, compromising accuracy.
Area of Science:
- Clinical chemistry
- Laboratory diagnostics
- Haematology
Background:
- A reference method for haemoglobin measurement, the haemiglobincyanide method, is available.
- A Dutch external quality assessment organization does not utilize this reference method for evaluating result trueness.
- This study investigates potential compromises in trueness due to the use of consensus values.
Purpose of the Study:
- To assess the trueness of haemoglobin measurements using Cell Dyn Sapphires.
- To compare the accuracy of Cell Dyn Sapphire results against the haemiglobincyanide reference method.
- To evaluate the reliability of consensus values from external quality assessment in determining haemoglobin measurement accuracy.
Main Methods:
- Haemoglobin concentration was measured on five Cell Dyn Sapphires across three locations.
- Results were compared to the established haemiglobincyanide reference method.
- Systematic deviations were analyzed using Passing and Bablok regression and Bland-Altman plots with patient and external quality assessment samples.
Main Results:
- Cell Dyn Sapphire haemoglobin measurements showed significant deviation from the reference method.
- Consensus results from external quality control samples also exhibited significant deviation from the reference method.
- A notable negative bias was observed in both the instrument measurements and the consensus values.
Conclusions:
- Cell Dyn Sapphires demonstrate a significant negative bias in haemoglobin measurement.
- External quality assessment consensus values display a greater negative bias compared to the reference method.
- External quality assessment should adopt the haemiglobincyanide reference method for improved trueness evaluation.
Related Concept Videos
Hemoglobin
When all four heme groups are bound to oxygen, the resulting molecule is called oxyhemoglobin. As a result, arterial blood...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Oxygen Transport in the Blood
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...