The relationship between scapular notching and reverse shoulder arthroplasty prosthesis design
Marc S Kowalsky1, Leesa M Galatz, Derek S Shia
1Department of Orthopaedic Surgery, Lenox Hill Hospital, Park Lenox Orthopaedics, New York, NY, USA.
Journal of Shoulder and Elbow Surgery
|November 15, 2011
Summary
The Tornier prosthesis showed a higher incidence of inferior scapular notching after reverse shoulder arthroplasty compared to the Zimmer prosthesis. Prosthesis design significantly impacts notching risk, influencing clinical outcomes.
Area of Science:
- Orthopedic surgery
- Biomedical engineering
- Radiology
Background:
- Inferior scapular notching is a frequent radiographic complication following reverse shoulder arthroplasty.
- Prosthesis design is a potential factor influencing the occurrence and severity of this complication.
Purpose of the Study:
- To evaluate the impact of different reverse shoulder arthroplasty prosthesis designs on the incidence and severity of inferior scapular notching.
- To compare notching rates between Tornier and Zimmer implant designs.
Main Methods:
- Retrospective review of 88 patients undergoing reverse shoulder arthroplasty with at least 12 months follow-up.
- Patients were grouped by prosthesis design: Tornier, Zimmer with nonretentive liner, and Zimmer with retentive liner.
- Notching was graded using the Sirveaux and a novel classification system.
Main Results:
- The Tornier prosthesis had a significantly higher incidence of notching (92%) compared to Zimmer designs (58%-71%).
- Tornier implants also showed significantly higher rates of high-grade notching and baseplate involvement.
- Retentive liners in Zimmer implants did not significantly alter notching rates compared to nonretentive liners.
Conclusions:
- The Tornier reverse arthroplasty demonstrated a higher incidence and severity of scapular notching than the Zimmer design at short-term follow-up.
- Differences in polyethylene morphology or glenosphere offset may contribute to the observed variations.
- Prosthesis design, alongside surgical technique, is crucial for mitigating scapular notching risk.