Effects of E-BEAM sterilization on drug-eluting stents: paclitaxel degradation elucidated by LC-MS-MS with
Yuri V Il'ichev1, Lori Alquier
1Cordis Corporation, Spring House, PA 19477, USA. iilitche@its.jnj.com
Abstract:
Effects of sterilization by electron beam (E-BEAM) on paclitaxel (1) mixed with poly(DL-lactide-co-glycolide) (PLG) in reservoirs of COSTAR Stents are examined by using liquid chromatography-mass spectrometry (LC-MS-MS) techniques with information-dependent acquisition (IDA). Numerous degradation products of 1 are formed in a β-radiation dose-dependent manner to give plethora of low-level degradants. This behavior, together with multiple interferences from PLG-related compounds, creates considerable challenges for analysis of the drug/PLG mixtures. IDA methods with different survey scans are proven to be very efficient in elucidating degradation pathways and in identifying numerous products. Combined LC-MS-MS results from analysis of sterilized drug substance and stents indicate that water addition and oxidative processes together with the isomerization are largely responsible for degradation of 1 under E-BEAM sterilization conditions used.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...