The acute effects of changes to AV delay on BP and stroke volume: potential implications for design of pacemaker
Charlotte H Manisty1, Ali Al-Hussaini, Beth Unsworth
1International Centre for Circulatory Health, Imperial College London, London, UK. cmanisty@ic.ac.uk
Background:
The AV delay optimization of biventricular pacemakers (cardiac resynchronization therapy) may maximize hemodynamic benefit but consumes specialist time to conduct echocardiographically. Noninvasive BP monitoring is a potentially automatable alternative, but it is unknown whether it gives the same information and similar precision (signal/noise ratio). Moreover, the immediate BP increment on optimization has been reported to decay away: it is unclear whether this is the result of an (undesirable) decrease in stroke volume or a (desirable) compensatory relief of peripheral vasoconstriction.
Methods And Results:
To discriminate between these alternative mechanisms, we measured simultaneous beat-to-beat stroke volume (flow) using Doppler echocardiography, and BP using finger photoplethysmography, during and after AV delay changes from 40 to 120 ms in 19 subjects with cardiac pacemakers. BP and stroke volume both increased immediately (P<0.001, within 1 heartbeat). BP showed a clear decline a few seconds later (average rate, -0.65 mm Hg/beat; r=0.95 [95% CI, 0.86-0.98]); in contrast, stroke volume did not decline (P=0.87). The immediate BP increment correlated strongly with the stroke volume increment (r=0.74, P<0.001). The signal/noise ratio was 3-fold better for BP than stroke volume (6.8±3.5 versus 2.3±1.4; P<0.001).
Conclusions:
Improving AV delay immediately increases BP, but the effect begins to decay within a few seconds. Reassuringly, this is because of compensatory vasodilatation rather than reduction in cardiac function. Pacemaker optimization will never be reliable unless there is an adequate signal/noise ratio. Using BP rather than Doppler minimizes noise. The early phase (before vascular compensation) has the richest signal lode.
Related Concept Videos
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Measurement of Blood Pressure
Pathophysiology of Cardiac Performance
Decreased pulse rate
There are specific risk factors that can elevate the likelihood of developing bradycardia. Advanced age is a significant factor, with bradycardia...
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...