Related Experiment Video
Updated: May 27, 2026

Analyzing the Photo-oxidation of 2-propanol at Indoor Air Level Concentrations Using Field Asymmetric Ion Mobility Spectrometry
Published on: June 14, 2018
[Determination of diborane in the air of workplace by ICP-AES]
Chun-Guang Ding1, Jing Zhang, Hui-Fang Yan
1National Institute of Occupational Health and Poison Control, Chinese Center for Disease Control, Beijing 100050, China.
Objective:
A sampling method was established to collect diborane in the air of workplace and an ICP-AES method was developed to determine the Boron in desorbed solution.
Method:
Diborane in the air of workplace was collected by solid sorbent tube filled with oxidant impregnated activated carbon. The adsorbed diborane was desorbed into 3% H2O2 aqueous, and then the desorbed Boron was determined by ICP-AES.
Results:
The sampling efficiency of this method was 99.6% with the desorption efficiency of diborane with 5.660 microg and 56.6 microg spiked were 90.9% and 99.5%, respectively. Both the intra-and inter-precision RSD were less than 8%. The standard curve of this method ranged from 0.1 to 10.0 microg/ml (Boron), and the LOD and LOQ were 0.011 mg/m3 and 0.035 mg/m3 (15L samples) respectively.
Conclusion:
The method established was suitable for diborane sampling and determination in the air of workplace.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Gas Chromatography: Types of Detectors-II