Related Experiment Video
Updated: May 27, 2026

Employing the Forced Oscillation Technique for the Assessment of Respiratory Mechanics in Adults
Published on: February 9, 2022
Physiologically acceptable resistance of an air purifying respirator
Barbara E Shykoff1, Dan E Warkander
1Navy Experimental Diving Unit, 321 Bullfinch Rd, Panama City, FL 32407, USA. barbara.shykoff.ca@navy.mil
Abstract:
Physiologically acceptable limits of inspiratory impediment for air purifying respirators (APRs) were sought.Measurements on 30 subjects included pressure in, and flow through, an APR, and respiratory and cardiovascular variables. Exercise with and without APR included ladder climbing, load lift and transfer, incremental running and endurance running, with endurance at 85% peak oxygen uptake. Resistance that did not alter minute ventilation (VE) was judged acceptable long-term. Acceptable short-term impediments were deduced from end exercise conditions. Proposed long-term limits are inspiratory work of breathing per tidal volume (WOBi/VT) ≤ 0.9 kPa and peak inspiratory pressure (P (i) peak) ≤1.2 kPa. Proposed short-term limits are: for VE ≤110 L min(-1), WOBi/VT ≤1.3 kPa and P (i) peak ≤ 1.8 kPa; and for VE >130 L min(-1), WOBi/VT ≤1.6 kPa. A design relation among VE, pressure–flow coefficients of an APR, and WOBi/VT is proposed. STATEMENT OF RELEVANCE: This work generalises results from one APR by considering the altered physiological parameters related to factors inhibiting exercise. Simple expressions are proposed to connect bench-test parameters to the relation between ventilation and work of breathing. Population-based recommendations recognise that those who need more air flow can also generate higher pressures.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Personal Protective Equipment
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.
The Functional Residual Capacity (FRC) represents the air in the...
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Respiratory Volumes and Capacities I
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: