The reliability and accuracy of a noncontact electrocardiograph system for screening purposes
Michael Czaplik1, Benjamin Eilebrecht, Rafael Walocha
1Department of Anesthesiology, University Hospital RWTH Aachen, Aachen, Germany. mczaplik@ukaachen.de
Background:
Electrocardiography (ECG) requires the application of electrodes to the skin and often necessitates undressing. Capacitively coupled electrodes embedded in a normal chair would be a rational alternative for ECG screening. We evaluated the reliability and accuracy of ECG electrodes imbedded in a chair cushion.
Methods:
Two independent clinicians compared ECG recordings obtained using skin electrodes with recordings obtained using capacitively coupled electrodes that were embedded in a chair cushion in an anesthesiology premedication room, a cardiology outpatient ward, and a cardiology day ward. We analyzed the data to compare the sensitivity and specificity for the diagnosis of cardiac arrhythmias.
Results:
ECG recordings were obtained from 107 patients. Heart rate was accurately measured using the capacitively coupled electrodes, but motion artifacts made the identification of P and T waves unreliable. Signal quality was poor for patients with low body weight, patients wearing clothing containing mixed fibers, and patients wearing sweaty shirts.
Conclusions:
Heart rate was accurately measured, and some cardiac arrhythmias were correctly diagnosed using capacitive ECG electrodes. Capacitive electrodes embedded into an examination chair are a promising tool for preoperative screening. Improved artifact reduction algorithms are needed before capacitive electrodes will replace skin electrodes.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin to...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Holter Monitor: 24-Hour Monitoring
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage. When...