Optimization of the CT acquisition protocol to reduce patient dose without compromising the diagnostic quality for
Sunil Kumar1, Anil Kumar Pandey, Punit Sharma
1Department of Nuclear Medicine, All India Institute of Medical Sciences, New Delhi, India.
Purpose:
We conducted a phantom study to optimize the computed tomography (CT) acquisition protocol, to reduce patient dose without compromising the diagnostic quality of PET-CT study.
Methods:
We conducted 42 PET-CT phantom studies. The jaszczak phantom was filled with 18F-fluorodeoxyglucose activity (1.4, 4, 6, 8, 10, and 12 mCi, on different settings) and PET-CT was acquired at different tube currents of 16, 25, 30, 35, 40, 45, and 50 mAs, keeping all the other parameters constant. Two nuclear medicine physicians qualitatively evaluated 252 images for their diagnostic quality (contrast, resolution, and number of lesion detected). The images were also evaluated qualitatively for noise on CT (the physical measure of image quality) and standardized uptake value (mean) on PET images. The weighted CT dose index (CTDIw) was measured for calculating the changes in the radiation dose with the change in tube current.
Results:
The 42 images provided same diagnostic information in CT images at different tube currents. However, there was a gradual increase in noise in the image acquired at lower values of the tube current, in mAs. The modulation of CT tube current had no effect on the PET images as well as the PET-CT images. Standardized uptake value mean values were also not affected with lowering of tube current. On reducing the tube current from 50 to 16 mAs, a reduction in the radiation dose (CTDIw) to almost 68% was achieved.
Conclusion:
It is possible to reduce the tube current of the CT tube of PET-CT to a much lower level, without any loss of diagnostic information of a PET-CT image. Moreover, this can significantly reduce the radiation dose.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...