Validation of plusoptiX S04 photoscreener as a vision screening tool in children with intellectual disability
Silay Canturk Ugurbas1, Atilla Alpay, Huseyin Tutar
1Zonguldak Karaelmas University, Department of Ophthalmology, Kozlu, Zonguldak, Turkey. silaycanturk@gmail.com
Purpose:
Intellectual disability (ID) is a major public health issue, affecting more than 1% of children worldwide. Pediatric vision screening using standard eye charts may be challenging for children with ID, who may benefit from a quick noninvasive vision screening tool. This study evaluated the accuracy of plusoptiX S04 (Plusoptix Gmbh, Nuremberg, Germany) in detecting amblyopia risk factors in children with ID.
Methods:
Children diagnosed with ID according to age-appropriate psychological tests were examined consecutively. Vision screening was performed using the plusoptiX S04 prior to complete ophthalmological examination, including cycloplegic refraction. Pass/refer screening results was compared with findings of ophthalmological examination. Amblyogenic risk factors were defined according to American Association for Pediatric Ophthalmology and Strabismus referral criteria.
Results:
A total of 182 infants and children were examined. Only 3% of children were uncooperative for screening. Ophthalmological examination detected amblyogenic risk factors in 32% of patients. The sensitivity of plusoptiX S04 was 95% (95% CI, 85.6%-98.9%), and specificity was 50% (95% CI, 40.8%-59.1%). The positive predictive value was 47% (95% CI, 37.7%-56.4%) and the negative predictive value was 95.4% (95% CI, 87.1%-99.1%).
Conclusions:
High sensitivity and negative predictive values suggest that the plusoptiX S04 will detect most children with ID who have amblyogenic risk factors, but one-half of all children referred will have no risk factors.