Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Bench studies evaluating devices for non-invasive ventilation: critical analysis and future perspectives
Carlo Olivieri1, Roberta Costa, Giorgio Conti
1Anesthesia and Intensive Care Medicine, Maggiore della Carità University Hospital, Novara, Italy.
Bench studies comparing non-invasive mechanical ventilation (NIV) devices lack consistency due to varied methods and terminology. Standardizing settings and criteria is crucial for reliable NIV device performance evaluation.
Area of Science:
- Respiratory Medicine
- Biomedical Engineering
- Medical Device Technology
Background:
- Non-invasive mechanical ventilation (NIV) is increasingly utilized in clinical practice.
- Numerous ventilators and interfaces for NIV have been introduced, necessitating performance evaluations.
- Bench studies comparing NIV devices are vital for informed equipment acquisition but often lack consistency.
Purpose of the Study:
- To review and discuss technical aspects influencing discrepancies in bench studies of NIV devices.
- To provide clinicians with valuable information on ventilators and interfaces for NIV.
- To identify factors contributing to inconsistent performance comparisons of NIV equipment.
Main Methods:
- Review of bench evaluation studies for NIV devices.
- Focus on specific technical aspects: lung models, inspiratory demand simulation, respiratory system properties, air leaks, ventilator modes, and interface performance.
- Analysis of methodologies for assessing trigger function, cycling, and pressurization rates.
Main Results:
- Lack of clear impact from different test lung models.
- Absence of standard references for simulated demand, air leaks, respiratory system properties, and ventilator settings.
- Variability in criteria for assessing trigger function, I:E cycling, pressurization rates, and inconsistent terminology.
Conclusions:
- Consistent experimental settings are needed to improve bench assessment of NIV devices.
- Uniform terminology and standard measurement criteria are essential for reliable performance comparisons.
- Standardization will enhance the evaluation of characteristics and performance of NIV ventilators and interfaces.
Related Concept Videos
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...