Related Experiment Video
Updated: May 27, 2026

An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
The primary stability of angle-stable versus conventional locked intramedullary nails
Dirk Wähnert1, Yves Stolarczyk, Konrad L Hoffmeier
1Department of Trauma-, Hand- and Reconstructive Surgery, Friedrich Schiller University Jena, Erlanger Allee 101, 07747, Jena, Germany. Dirk.Waehnert@ukmuenster.de
Purpose:
The aim of this study was to compare the initial biomechanical characteristics of the angle-stable locking system for intramedullary nails using the new biodegradable sleeve with conventional locking in the treatment of unstable distal tibial fractures.
Methods:
Eight pairs of fresh, frozen porcine tibiae were used for this study. The expert tibial nail (Synthes) was equipped with either conventional locking screws (CL) or the angle-stable locking system (AS). This system consists of a special ASLS screw with a biodegradable sleeve. For this investigation distal tibias (5.5 cm) were used and the nails were locked with three screws in both groups. Biomechanical testing included non-destructive torsional and axial loading.
Results:
The AS group showed a significantly higher torsional stiffness (70%) compared to the CL group. The range of motion was 0.5 times smaller for the AS constructs. The neutral zone was eight times higher in the CL group (p < 0.001). In axial loading the AS group also showed a 10% higher axial stiffness and a 12% lower range of motion (p < 0.001).
Conclusion:
The angle-stable locking system (ASLS) using a special screw and sleeve locking for intramedullary nails provides a significantly higher primary stability. The differences determined in this study may have clinical relevance particularly for torsional loads. For the new biodegradable angle-stable sleeve we found a comparable stability to the PEEK-based sleeve system. This system has the potential to decrease complications such as secondary loss of reduction and mal-/non-union.
More Related Videos
04:41Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Stability of structures
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Wedges
Consider using a wedge to lift a heavy slab. Here, the wedge functions by converting the applied force into a much larger force directed almost perpendicular to the initial force. This...