Do bulb syringes conform to neonatal resuscitation guidelines?
Pradeep Alur1, Jonathan Liss, Frank Ferrentino
1Department of Pediatrics, York Hospital, 1001 S. George Street, York, PA 17405, USA. palur@wellspan.org
Resuscitation
|December 7, 2011
Summary
Most neonatal suction bulbs generate pressures exceeding safety guidelines, risking airway injury. Only one bulb met the American Academy of Pediatrics Neonatal Resuscitation Program recommendations, highlighting a need for awareness among healthcare providers.
Area of Science:
- Neonatal care
- Medical device safety
- Pediatric airway management
Background:
- High suction pressures during neonatal resuscitation can cause airway injury.
- The American Academy of Pediatrics Neonatal Resuscitation Program (NRP) recommends suction pressures below 100 mm Hg.
- The safety and efficacy of commercially available suction bulbs in meeting these guidelines are not well-established.
Purpose of the Study:
- To evaluate whether commercially available suction bulbs conform to the NRP's recommended suction pressure limits.
- To assess the pressures generated by standard neonatal suction bulbs during simulated resuscitation procedures.
Main Methods:
- A prospective in vitro study was conducted.
- Twenty-five personnel involved in neonatal resuscitation used their standard techniques to squeeze six different commercially available suction bulbs.
- A calibrated pneumatic transducer measured the pressure generated by each squeeze.
Main Results:
- Only one of the six tested suction bulbs met the NRP guideline of generating pressures less than 100 mm Hg.
- All participants exceeded the 100 mm Hg threshold with at least one bulb type (p<0.001).
Conclusions:
- The majority of neonatal suction bulbs tested do not comply with NRP guidelines for safe suction pressures.
- The one bulb that met guidelines was a larger size (3 oz), potentially limiting its use in premature infants.
- Healthcare providers must be aware of the pressures generated by suction bulbs to prevent iatrogenic airway injury in newborns.