Related Experiment Video
Updated: May 26, 2026

Treating Low Back Pain in Failed Back Surgery Patients with Multicolumn-lead Spinal Cord Stimulation
Published on: June 26, 2018
Challenges to setting spinal cord stimulator parameters during intraoperative testing: factors affecting coverage of
Assaf T Gordon1, Sheng Ping Zou, Yong Kim
1Departments of Physical Medicine and Rehabilitation, Anesthesiology, Pain Management Center, and Orthopedic Surgery, New York University Medical Center, New York, USA.
Abstract:
Objective. Spinal cord stimulator (SCS) parameter settings have been well studied; however, the goal of this exploratory study was to examine the SCS parameters used during intra-operative stimulation (IOS) at trial lead placement. Methods. In this retrospective study, we report the IOS parameter settings for 22 patients who underwent thoracic SCS lead trial for treatment of refractory low back and/or leg pain. Results. Paresthesia coverage was shown to differ depending upon the pain syndrome and the region involved (back and/or leg, p = 0.03). Certain stimulation parameters were demonstrated to be linked, including pulse width with rate (p = 0.04) and bipolar activation distance with amplitude (p < 0.01). Important variations in field configuration practice patterns also emerged. Conclusions. Larger prospective studies are required to confirm and extend the current results. The ultimate goal for this report is to establish a foundation for future studies to create an evidence-based standardized algorithm for IOS to enhance the success rate of SCS trial screening.