Type and accuracy of sphygmomanometers in primary care: a cross-sectional observational study
Christine A'Court1, Richard Stevens, Sarah Sanders
1Department of Primary Health Care, University of Oxford, Oxford. christine.acourt@phc.ox.ac.uk
Background:
Previous studies identified worrying levels of sphygmomanometer inaccuracy and have not been repeated in the era of digital measurement of blood pressure.
Aim:
To establish the type and accuracy of sphygmomanometers in current use.
Design And Setting:
Cross-sectional, observational study in 38 Oxfordshire primary care practices.
Method:
Sphygmomanometers were evaluated between 50 and 250 mmHg, using Omron PA350 or Scandmed 950831-2 pressure meters.
Results:
Six hundred and four sphygmomanometers were identified: 323 digital (53%), 192 aneroid (32%), 79 mercury (13%), and 10 hybrid (2%) devices. Of these, 584 (97%) could be fully tested. Overall, 503/584 (86%) were within 3 mmHg of the reference, 77/584 (13%) had one or more error of 4-9 mmHg, and 4/584 (<1%) had one or more error of more than 10 mmHg. Mercury (71/75, 95%) and digital (272/308, 88%) devices were more likely to be within 3 mmHg of the reference standard than aneroid models (150/191, 78%) (Fisher's exact test P = 0.001). Donated aneroid devices from the pharmaceutical industry performed significantly worse: 10/23 (43%) within 3 mmHg of standard compared to 140/168 (83%) aneroid models from recognised manufacturers (Fisher's exact test P<0.001). No significant difference was found in performance between manufacturers within each device type, for either aneroid (Fisher's exact test P = 0.96) or digital (Fisher's exact test P = 0.7) devices.
Conclusion:
Digital sphygmomanometers have largely replaced mercury models in primary care and have equivalent accuracy. Aneroid devices have higher failure rates than other device types; this appears to be largely accounted for by models from indiscernible manufacturers. Given the availability of inexpensive and accurate digital models, GPs could consider replacing aneroid devices with digital equivalents, especially for home visiting.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...