Related Experiment Video
Updated: May 26, 2026

Inspiratory Muscle Training as an Adjunct to the Treatment of Weaning Failure in Critically Ill Patients: A Practical Guide
Published on: January 30, 2026
Potential inadequacy of automatic tube compensation to decrease inspiratory work load after at least 48 hours of
Jun Oto1, Hideaki Imanaka, Emiko Nakataki
1Department of Emergency and Critical Care Medicine, The University of Tokushima Graduate School, Tokushima City, Japan. joto@clin.med.tokushima-u.ac.jp
Background:
Endotracheal tubes (ETTs) impose a substantial respiratory work load on mechanically ventilated patients. Automatic tube compensation (ATC) should overcome this flow-dependent resistive work load; however, ETT resistance can be increased by tracheal secretions or ETT deformities. Our bench study investigated whether ATC provides effective respiratory work load relief in used ETTs.
Methods:
We enrolled 20 critically ill patients requiring mechanical ventilation for longer than 48 hours. After extubation, we collected the used ETTs and measured the pressure-time products (PTPs) by using a bellows-in-a-box lung model that simulated spontaneous breathing, at a respiratory rate of 10 breaths/min, inspiratory time of 1.0 s, and tidal volumes (V(T)) of 300 mL, 500 mL, and 700 mL. The ventilator was set at ATC 100% with PEEP of 5 cm H(2)O and F(IO(2)) of 0.21. The flow and airway pressure at the proximal (P(aw)) and distal (P(tr)) ends of the ETT were recorded, and the PTP integrated from P(tr) analyzed.
Results:
PTP values increased with V(T) during ATC. Even at 100% ATC the ventilator did not completely compensate for the PTP imposed by the ETT. In used ETTs, peak flow and peak P(aw) were lower and PTP values were greater than in new ETTs. As V(T) increased, the percentage difference in the PTP values between the new and used ETTs increased.
Conclusions:
ATC does not necessarily compensate for an ETT-imposed respiratory work load. ETT configuration changes and tracheal secretions can increase ETT resistance and decrease the ability of ATC to compensate for the increased respiratory work load.
Related Concept Videos
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Atelectasis II: Pathophysiology