Use of Carr-Purcell pulse sequence with low refocusing flip angle to measure T₁ and T₂ in a single experiment
Fabiana Diuk De Andrade1, Antonio Marchi Netto, Luiz Alberto Colnago
1Instituto de Química de São Carlos, Universidade de São Paulo, Av. Trabalhador São-carlense 400, São Carlos, SP 13560-970, Brazil. fabianadiuk@yahoo.com.br
Abstract:
The Carr-Purcell pulse sequence, with low refocusing flip angle, produces echoes midway between refocusing pulses that decay to a minimum value dependent on T(2). When the refocusing flip angle was π/2 (CP(90)) and τ>T(2), the signal after the minimum value, increased to reach a steady-state free precession regime (SSFP), composed of a free induction decay signal after each pulse and an echo, before the next pulse. When τ