Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
A double faced ionization chamber for quality control in diagnostic radiology beams
Jonas O Silva1, Linda V E Caldas
1Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN/SP), Av. Prof. Lineu Prestes, 2242, 05508-000, São Paulo, Brazil. josofisico@yahoo.com.br
Abstract:
The development of new radiation detectors of low cost but with adequate materials is a very important task for countries that have to import ionization chambers such as Brazil. A special double faced ionization chamber was developed for use in conventional diagnostic radiology beams and computed tomography energy ranges. The results show that this new chamber present applicability in conventional diagnostic radiology and computed tomography quality control programs.