The fixation strength of scaphoid bone screws: an in vitro investigation using polyurethane foam
Louise A Crawford1, Eric S Powell, Ian A Trail
1Centre for Hand and Upper Limb Surgery, Wigan, Lancashire, UK.
Purpose:
To compare the compression strength and pull-apart resistance of 5 single-piece scaphoid bone compression screws (Acutrak, Asnis, Herbert, Herbert-Whipple, and Little Grafter), with those of 2 dual-component screws (Kompressor and TwinFix).
Methods:
Two blocks of polyurethane foam were compressed with a screw while held in a tension test machine, with the force measured at full insertion of the screw. The 2 blocks were then pulled apart, and the maximum resistive force was measured.
Results:
The dual-component screws (Kompressor and TwinFix) gave greater compression force than the single-component screws, with the Kompressor screw giving statistically significantly greater compression than the TwinFix. The pull-apart resistance forces did not show such clear differences.
Conclusions:
The Kompressor screw achieves the greatest compressive forces and has one of the highest pull-apart forces.
Clinical Relevance:
When compression and pull-apart resistance are considered, the Kompressor screw has advantages over other methods of scaphoid fixation.