Evaluation of 25-hydroxy vitamin D assay on the immunodiagnostic systems iSYS analyser
Zee N Cluse1, Andrew N Fudge, Malcolm J Whiting
1Metabolic Laboratory, Chemical Pathology Directorate, SA Pathology, Box 14, Rundle Post Office, Adelaide SA 5000, Australia. zee.cluse@health.sa.gov.au
Background:
We evaluated the recently released chemiluminescence assay for 25-hydroxy vitamin D (25-OHD) on the Immunodiagnostic Systems iSYS (IDS-iSYS) automated analyser.
Methods:
The IDS-iSYS comparison was performed using patient samples previously measured for 25-OHD by a validated liquid chromatography-tandem mass spectrometry (LC-MS/MS) method (n = 119) and an IDS enzyme immunoassy (IDS-EIA) method (n = 64). Limit of detection and limit of quantification were determined from a precision profile. Imprecision was assessed using quality control material and pooled serum. External QAP material (Vitamin D External Quality Assessment Scheme, UK) was analysed to establish inaccuracy. Linearity was assessed by two dilution studies. Cross-reactivity was determined by three serial dilution studies of patient samples with known 25-OHD(2) concentrations.
Results:
The IDS-iSYS correlated well with both established methods (iSYS = 1.03LC-MS/MS - 6.53, R(2) = 0.82 and iSYS = 1.07IDS-EIA - 1.61, R(2) = 0.86). Imprecision of the iSYS assay for IDS control material was 13.4% at 32 nmol/L, 10% at 78 nmol/L, 9.4% at 161 nmol/L, and for the pooled material 9.3% at 72 nmol/L and 5.6% at 158 nmol/L. The evaluation found the assay to be highly accurate (IDS-iSYS = 0.93ALTM + 3.79, R(2) = 0.94) and linear (obs(1) = 0.93exp(1) - 5.05, R(2) = 0.99 (P = 0.256); and obs(2) = 0.97exp(2) + 6.07, R(2) = 0.97 (P = 0.654); ALTM, all-laboratory trimmed mean). Cross-reactivity studies demonstrated no significant difference to the calculated total 25-OHD as measured by LC-MS/MS.
Conclusions:
Even though the imprecision of the iSYS was found to be greater than that of the LC-MS/MS and EIA methods, the performance characteristics of the IDS-iSYS 25-OHD assay are suitable for routine diagnostic purposes on a high throughput automated analyser.
More Related Videos
08:48Dynamic Monitoring of Seroconversion using a Multianalyte Immunobead Assay for Covid-19
Published on: February 16, 2022
08:04Development of an IFN-γ ELISpot Assay to Assess Varicella-Zoster Virus-specific Cell-mediated Immunity Following Umbilical Cord Blood Transplantation
Published on: July 9, 2014
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Automated Microbial Diagnostics