Biomechanical comparison of occiput-C1-C2 fixation techniques: C0-C1 transarticular screw and direct occiput condyle
Tomoyuki Takigawa1, Peter Simon, Alejandro A Espinoza Orías
1Department of Orthopedic Surgery, Rush University Medical Center, Chicago, IL 60612, USA.
Study Design:
In vitro human cadaveric biomechanical study.
Objective:
The objective was to evaluate and compare the construct stability of occiput-C1-C2 fixation provided by C0-C1 transarticular screws or occipital condyle screws.
Summary Of Background Data:
The placement of an occipital plate is commonly recommended in occipitocervical fixation surgery. However, there are unique clinical situations in which the placement of the occipital plate may not be possible or may have already failed. For these situations, 2 novel techniques that use the occipital condyle have been recently introduced: (1) C0-C1 transarticular screws fixation and (2) direct occipital condyle screws and C1 lateral mass screws fixation. However, there is a lack of thorough biomechanics studies of these techniques.
Methods:
Nondestructive kinematic tests and destructive tests were conducted in 16 fresh frozen cadaveric spines. As a nondestructive kinematic test, a pure moment of up to 2.0 N·m was applied in smooth continuous flexion/extension, lateral bending, and axial rotation motions. In addition to an intact-state case, a total of 5 different constructs (standard occipital plate, C0-C1 transarticular screws with/without occipital plate, and occipital condyle screws with/without occipital plate) were tested after destabilization of C0-C1 and C1-C2. All constructs had C2 pedicle screws fixation, and occipital condyle screws were incorporated with C1 lateral screws.
Results:
All fixation techniques significantly reduced ranges of motion compared with the intact state. In comparison with the standard occipital plate construct, the 2 novel techniques showed higher stability in axial rotation and lower stability in lateral bending. In both nondestructive and destructive tests, there were no statistical differences between C0-C1 transarticular screw construct and occipital condyle screw construct.
Conclusion:
The C0-C1 transarticular screw technique and direct occipital condyle screw with C1 lateral mass screw technique can be salvage fixation methods when occipital plate fixation is not feasible.
Related Concept Videos
Cranial Bones: Superior and Posterior View
The frontal bone is the single bone that forms the forehead. At its anterior midline, between the eyebrows, there is a slight depression called the glabella. The frontal bone also forms the supraorbital margin of the orbit. Near the middle of this margin is the supraorbital foramen, the opening that provides passage for a sensory nerve to the forehead. The frontal bone is thickened just above each supraorbital margin,...
Sutures of the Skull
Sutures are immobile joints between adjacent bones of the skull. The narrow gap between the bones is filled with dense, fibrous connective tissue that unites the bones. The long sutures located between the skull bones are not straight but instead follow irregular, tightly twisting paths. These twisting lines tightly...