Real-time elastography of hepatic thermal lesions in vitro: histopathological correlation
P Wiggermann1, E-M Jung, S Glöckner
1Institut für Röntgendiagnostik, Klinikum der Universität Regensburg, Germany. philipp.wiggermann@klinik.uniregensburg.de
Purpose:
To evaluate the reliability of elastography, a new ultrasonographic method, for delineating thermal lesion boundaries in porcine liver tissue by comparing lesion dimensions determined by real-time elastography with the findings at gross pathology.
Materials And Methods:
A total of 15 thermal lesions with diameters ranging from 17 to 60 mm were created using radiofrequency ablation (RFA). Color-coded elastography was performed by one experienced examiner, using a 6 - 15 MHz high frequency linear transducer (LOGIQ E9, GE). Lesions were examined using B-mode and real-time elastography (RTE). Lesion detection, delineation and size were assessed using B-mode and RTE immediately after each thermal ablation ( < 5 min). Measurements of the sections representing the same image plane used for elastography were taken during pathologic examination and compared to the measurements obtained from the elastograms.
Results:
In our sample a statistically significant correlation in vitro between RTE and pathological measurements with respect to the lesion's principal axis and area (r2 = 0.9338 long axis, r2 = 0.8998 short axis and r2 = 0.9676 area) was found. Overall, elastography slightly underestimated the lesion size, as judged by the digitalized pathologic images.
Conclusion:
These results support that RTE outperforms conventional B-mode ultrasound and could potentially be used for the routine assessment of thermal therapies.
More Related Videos
07:13Application of Ultrasound and Shear Wave Elastography Imaging in a Rat Model of NAFLD/NASH
Published on: April 20, 2021
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Liver Histology
Hepatocytes perform a variety of essential functions. They secrete...