Left-ventricular outflow tract ventricular-tachycardia event following CoreValve transcatheter aortic-valve
Danny Dvir1, Moti Haim, Ran Kornowski
1Department of Cardiology, Rabin Medical Center, Petach Tikva, Israel.
Abstract:
We describe a patient with severe aortic stenosis who underwent CoreValve (Medtronic) implantation and presented several days later with a tachyarrhythmic episode. The electrocardiographic features of the arrhythmia were compatible with left-ventricular outflow tract ventricular-tachycardia. The life-threatening event was not associated with ischemia or an electrolyte disorder and was not drug-induced. A probable cause was the irritation of the myocardium by the prosthetic valve. Clinicians should be aware that the presence of anatomical parameters that increases the risk for myocardial injury and the need for pacemaker implantation might indicate an increased risk of a tachyarrhythmic episode.
Related Concept Videos
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization
Aortic Regurgitation I: Introduction
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...