Comparative in vitro validation of VistaProof and DIAGNOdent pen for occlusal caries detection in permanent teeth
Kyriaki Seremidi1, Panagiotis Lagouvardos, Katerina Kavvadia
1Department of Pediatric Dentistry, School of Dentistry, University of Athens, Greece.
Purpose:
Current caries diagnostic tools are neither very accurate nor very reliable for the detection of carious lesions of different depths. Thus, the development of new devices and techniques is needed. The aim of this in vitro study was to validate a newer fluorescence device VistaProof (VP), and compare it with DIAGNOdent Pen (DP), direct visual (DV) and indirect visual methods (IDV), with respect to accuracy and reliability for the detection of occlusal caries in permanent teeth.
Methods And Materials:
One hundred seven sites on 41 occlusal surfaces of recently extracted premolars were selected and classified into lesion categories according to Ekstrand's clinical criteria, by direct and indirect visual examination. The fluorescence of the sites was also measured by the two devices, and the teeth were ground through the sites for histological evaluation of their lesion depth. One calibrated examiner of high reliability (intraclass correlation coefficient [ICC]>0.85) made all of the evaluations. Sensitivity, specificity, and accuracy of each detection method were estimated based on histological examination as the reference method, estimated using cutoff limits calculated on the basis of best agreement between the devices' values and histological examination. McNemar tests and receiver operating characteristic (ROC) curve analyses were used to compare the validity measures of all detection methods at α=0.05, while the ICC was used to test the reproducibility of the methods based on a second measurement one week after the first.
Results:
There was no statistically significant difference (p>0.05) between the accuracy of DP and VPs for both enamel and dentin lesions. The areas under the ROC curves (AUC) for the two devices were also found not to be different (p>0.05). The reliability of DP was statistically significantly better than VP (p<0.05).
Conclusion:
The validity of both fluorescence devices were not found to be significantly different and not better than visual methods for the detection of noncavitated carious lesions.
More Related Videos
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
03:10Pilot In Vitro Study to Assess Cleaning Ability and Effects of Different Decontamination Methods on Implant Surfaces
Published on: November 21, 2025
Related Concept Videos
Assessment of the Mouth
Mouth Inspection
The inspection begins with visually examining the mouth for symmetry, color, and size.
Drug Product Performance: In Vitro–In Vivo Correlation