Related Experiment Video
Updated: May 26, 2026

Human Fetal Blood Flow Quantification with Magnetic Resonance Imaging and Motion Compensation
Published on: January 7, 2021
Passive fetal monitoring by adaptive wavelet denoising method
Sergey Vaisman1, Shimrit Yaniv Salem, Gershon Holcberg
1Department of Biomedical Engineering, Ben Gurion University of the Negev, Beer Sheva, Israel.
Abstract:
Fetal Heart Rate (FHR) monitoring is one of the most important fetal well being tests. Existing FHR monitoring methods are based on Doppler ultrasound technique, which has several disadvantages. Passive fetal monitoring by phonocardiography is an appropriate alternative; however, its implementation is a challenging task due to low energy of fetal heart sounds and multiple interference signals presence. In this paper, an advanced signal processing method for passive fetal monitoring based on adaptive wavelet denoising is presented. The method's performance is compared with Doppler ultrasound monitor. The results show 94-97.5% accuracy, including highly disturbed cases.