Related Experiment Video
Updated: May 26, 2026

Serum and Plasma Copy Number Detection Using Real-time PCR
Published on: December 15, 2017
Accuracy in copy number calling by qPCR and PRT: a matter of DNA
Nora Fernandez-Jimenez1, Ainara Castellanos-Rubio, Leticia Plaza-Izurieta
1Immunogenetics Research Laboratory, Cruces University Hospital, Barakaldo, Basque Country, Spain.
Abstract:
The possible implication of copy number variation (CNV) in the genetic susceptibility to human disease needs to be assessed using robust methods that can be applied at a population scale. In this report, we analyze the performance of the two major techniques, quantitative PCR (qPCR) and paralog ratio test (PRT), and investigate the influence of input DNA amount and template integrity on the reliability of both methods. Analysis of three genes (PRELID1, SYNPO and DEFB4) in a large sample set showed that both methods are prone to false copy number assignments if sufficient attention is not paid to DNA concentration and quality. Accurate normalization of samples is essential for reproducible qPCR because it avoids the effect of differential amplification efficiencies between target and control assays, whereas PRT is generally more sensitive to template degradation due to the fact that longer amplicons are usually needed to optimize sensitivity and specificity of paralog sequence PCR. The use of normalized, high quality genomic DNA yields comparable results with both methods.
Related Concept Videos
Comparing Copy Number Variations and SNPs
Copy number variations or CNVs are the structural variations that cover more than 1kb of DNA sequence. The single nucleotide polymorphism (SNP), on the other hand, is a single nucleotide change or a point mutation that is found in more than 1%...
Real Time RT-PCR
The real-time quantification of the number of amplified products is...
PCR
Genome Copying Errors