Related Experiment Video
Updated: May 26, 2026

Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Acceptability of virtual unenhanced CT of the aorta as a replacement for the conventional unenhanced phase
N Shaida1, D J Bowden, T Barrett
1Department of Radiology, Addenbrooke's Hospital, Cambridge, UK. nadeem.shaida@addenbrookes.nhs.uk
Aim:
To evaluate whether virtual unenhanced (VU) computed tomography (CT) images generated of the aorta were of sufficient quality to replace the conventional unenhanced (CU) images.
Materials And Methods:
Forty-nine patients undergoing examination of the thoracic or abdominal aorta were examined using a dual-energy protocol. VU images were generated from the arterial phase images and compared to the CU images. Objective analysis was performed by drawing paired regions of interest (ROIs) within the thoracic and abdominal aorta and measuring the radiodensity in Hounsfield units attenuation within the ROIs. Subjective analysis was performed by two experienced readers evaluating the VU images in terms of noise, quality, calcium loss, and overall acceptability.
Results:
The attenuation was significantly higher in the VU images compared to the CU images within the thoracic aorta (p < 0.01) but not within the abdominal aorta (p = 0.15). Overall the VU images of the abdominal aorta were deemed acceptable as replacements for the CU images in 93% of cases. For the thoracic aorta, the VU images were deemed acceptable in only 12% of cases, primarily due to pulsation artefact.
Conclusion:
VU images of the abdominal aorta are acceptable as replacements for the CU images in the vast majority of cases; however, they are not suitable as replacements for the CU images of the thoracic aorta.