Related Experiment Video
Updated: May 26, 2026

Electromagnetic Source Imaging in Presurgical Evaluation of Children with Drug-Resistant Epilepsy
Published on: September 20, 2024
Ictal MEG onset source localization compared to intracranial EEG and outcome: improved epilepsy presurgical
Hisako Fujiwara1, Hansel M Greiner, Nat Hemasilpin
1Division of Neurology, Cincinnati Children's Hospital Medical Center, Cincinnati, OH, USA. Hisako.Fujiwara@cchmc.org
Purpose:
Magnetoencephalography (MEG) has been shown a useful diagnostic tool for presurgical evaluation of pediatric medically intractable partial epilepsy as MEG source localization has been shown to improve the likelihood of seizure onset zone (SOZ) sampling during subsequent evaluation with intracranial EEG (ICEEG). We investigated whether ictal MEG onset source localization further improves results of interictal MEG in defining the SOZ.
Methods:
We identified 20 pediatric patients with one habitual seizure during MEG recordings between October 2007 and April 2011. MEG was recorded with sampling rates of 600Hz and 4000Hz for 10 and 2min respectively. Continuous head localization (CHL) was applied. Source localization analyses were applied using multiple algorithms, both at the beginning of ictal onset and for interictal MEG discharges. Ictal MEG onsets were identified by visual inspection and power spectrum using short-time Fourier transform (STFT). Source localizations were compared with ICEEG, surgical procedure and outcome.
Key Findings:
Eight patients met all inclusion criteria. Five of the 8 patients (63%) had concordant ictal MEG onset source localization and interictal MEG discharge source localizations in the same lobe, but the source of ictal MEG onset was closer to the SOZ defined by ICEEG.
Significance:
Although the capture of seizures during MEG recording is challenging, the source localization for ictal MEG onset proved to be a useful tool for presurgical evaluation in our pediatric population with medically intractable epilepsy.