Related Experiment Video
Updated: May 26, 2026

Characterization of Recombination Effects in a Liquid Ionization Chamber Used for the Dosimetry of a Radiosurgical Accelerator
Published on: May 9, 2014
Characterization of an extrapolation chamber for low-energy X-rays: experimental and Monte Carlo preliminary results
Lucio P Neves1, Eric A B Silva, Ana P Perini
1Instituto de Pesquisas Energéticas e Nucleares (IPEN-CNEN), Comissão Nacional de Energia Nuclear, Av. Prof. Lineu Prestes 2242, 05508-000 São Paulo, SP, Brazil. lpneves@ipen.br
Abstract:
The extrapolation chamber is a parallel-plate ionization chamber that allows variation of its air-cavity volume. In this work, an experimental study and MCNP-4C Monte Carlo code simulations of an ionization chamber designed and constructed at the Calibration Laboratory at IPEN to be used as a secondary dosimetry standard for low-energy X-rays are reported. The results obtained were within the international recommendations, and the simulations showed that the components of the extrapolation chamber may influence its response up to 11.0%.