Related Experiment Video
Updated: May 26, 2026

Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
Corrective osteotomy of distal radial deformities: a new method of guided locking fixed screw positioning
1Balgrist University Hospital, University of Zürich, Zürich, Switzerland. Mazda.Farshad@balgrist.ch
Abstract:
Corrective osteotomy for distal radial malunion is a valuable but at times technically challenging operation. We have developed a new device to aid in the performance of the operation. We compared clinical use of the new technique with the standard technique. In 11 patients treated with the new technique the volar locking plate needed repositioning only once. With the standard technique in 17 patients the plate was repositioned in nine cases and needed bending in six cases. The new method corrected radial inclination (SD of 3° vs. 9°) and ulnar variance (SD of 0.9 mm vs. 1.4 mm) more predictably than the standard technique but there was no advantage in correction of volar tilt (SD of 6° vs. 4°). In our experience the new device makes corrective osteotomy of the distal radius easier and more reliable.
More Related Videos
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
11:21Development of a Novel Internal Fixation Model for Rat Radial Fractures: Fracture Healing Assessment and Dorsal Root Ganglion Isolation
Published on: March 13, 2026