Phase-specific cone beam computed tomography reduces reconstructed volume loss of moving phantom
1Division of Radiation Oncology, Department of Oncology, National Taiwan University Hospital, No. 7, Chung-Shan South Road, 10002, Taipei, Taiwan.
Purpose:
The accurate volumetric calculation of moving targets/organs is required to use cone-beam computed tomography (CBCT) for replanning purposes. This study was aimed to correct the reconstructed volume losses of moving phantoms by phase-specific CBCT.
Materials And Methods:
Planning fan-beam CT (FBCT) of five hepatobiliary/gastrointestinal/pancreatic cancer patients were acquired under active breathing control and compared with free-breathing CBCT for kidney volumes. Three different-sized ball phantoms were scanned by FBCT and CBCT. Images were imported to a planning system to compare the reconstructed volumes. The phantoms were moved longitudinally on an oscillator with different amplitudes/frequencies. The phase-specific projections of CBCT for moving phantoms were selected for volume reconstruction.
Results:
The differences in reconstructed volumes of static small, medium, large phantoms between FBCT and CBCT were - 6.7%, - 2.3%, and - 2.0%, respectively. With amplitudes of 7.5-20 mm and frequencies of 8-16 oscillations/min, volume losses on CBCT were comparable with FBCT in large moving phantoms (range 9.1-27.2%). Amplitudes were more subject to volume losses than frequencies. On phase-specific CBCT, volume losses were reduced to 2.3-6.5% by reconstruction using 2-3 projections at end/midoscillation phase.
Conclusion:
Amplitude had more impact than frequency on volume losses of moving phantoms on CBCT. Phase-specific CBCT reduced volume losses.
More Related Videos
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography