Related Experiment Video
Updated: May 26, 2026

Using an Ingestible Telemetric Temperature Pill to Assess Gastrointestinal Temperature During Exercise
Published on: October 7, 2015
Validity of field expedient devices to assess core temperature during exercise in the cold
James R Bagley1, Daniel A Judelson, Barry A Spiering
1Exercise Physiology Laboratory, Department of Kinesiology, California State University, Fullerton, CA, USA.
Introduction:
Exposure to cold environments affects human performance and physiological function. Major medical organizations recommend rectal temperature (TREC) to evaluate core body temperature (TcORE) during exercise in the cold; however, other field expedient devices claim to measure TCORE. The purpose of this study was to determine if field expedient devices provide valid measures of TcRE during rest and exercise in the cold.
Methods:
Participants included 13 men and 12 women (age = 24 +/- 3 yr, height = 170.7 +/- 10.6 cm, mass = 73.4 +/- 16.7 kg, body fat = 18 +/- 7%) who reported being healthy and at least recreationally active. During 150 min of cold exposure, subjects sequentially rested for 30 min, cycled for 90 min (heart rate = 120-140 bpm), and rested for an additional 30 min. Investigators compared aural (T(AUR)), expensive axillary (T(AXLe)), inexpensive axillary (T(AXLi)), forehead (T(FOR)), gastrointestinal (T(GI)), expensive oral (T(ORLe)), inexpensive oral (T(ORLi)), and temporal (T(TEM)) temperatures to T(REc) every 15 min. Researchers used mean difference between each device and T(REC) (i.e., mean bias) as the primary criterion for validity.
Results:
T(AUR), T(AXLe), T(AXLi), T(FOR), TORLe, T(ORLi), and TTEM provided significantly lower measures compared to T(REC) and fell below our validity criterion. T(GI) significantly exceeded T(REC) at three of eleven time points, but no significant difference existed between mean T(REC) and T(GI) across time. Only T(GI) achieved our validity criterion and compared favorably to T(REC).
Conclusion:
T(GI) offers a valid measurement with which to assess T(CORE) during rest and exercise in the cold; athletic trainers, mountain rescuers, and military medical personnel should avoid other field expedient devices in similar conditions.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...