In vivo temperature measurement in the subacromial bursa during arthroscopic subacromial decompression
Scott L Barker1, Alan J Johnstone, Kapil Kumar
1Department of Orthopaedics, Woodend Hospital, Aberdeen, Scotland, UK. scott.barker@nhs.net
Background:
The purpose of the study was to evaluate whether use of a bi-polar radiofrequency (RF) ablation wand would cause excess heating, which may lead to collateral damage to the surrounding tissues during arthroscopic subacromial decompression. Cadaveric studies have shown that high temperatures can potentially be reached when using RF ablation wands in arthroscopic shoulder surgery. Only 1 other published study assesses these temperature rises in the clinical setting.
Methods:
Fifteen patients were recruited to participate in the study. A standard arthroscopic subacromial decompression was performed using continuous flow irrigation, with intermittent use of the RF ablation wand for soft tissue debridement. The temperature of the irrigation fluid within the subacromial bursa and the outflow fluid from the suction port of the wand were measured during the procedure using fiber-optic thermometers.
Results:
The mean peak temperature recorded in the subacromial bursa was 32.0°C (29.3-43.1°C), with a mean rise from baseline of 9.8°C. The mean peak temperature recorded from the outflow fluid from the wand was 71.6°C (65.6-77.6°C), with a mean rise from baseline of 49.4°C.
Conclusion:
High temperatures were noted in the outflow fluid from the wand; however, this was not evident in the subacromial bursa itself. Use of room temperature inflow fluid, maintenance of flow through the bursa, and avoidance of prolonged uninterrupted use of the wand all appear to ensure that safe temperatures are maintained in the subacromial bursa not only in the laboratory but also in a clinical setting.
Related Concept Videos
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...