Disparity in left ventricular stimulation among different pacing configurations in cardiac resynchronization therapy
Jose F Huizar1, Karoly Kaszala, Jayanthi Koneru
1McGuire VA Medical Center and Virginia Commonwealth University/Pauley Heart Center, Richmond, VA 23249, USA. jose.huizar2@va.gov
Background:
Cardiac resynchronization therapy (CRT) demands high energy utilization due to continuous biventricular pacing. Current technology allows 6 pacing configurations for a bipolar left ventricular (LV) lead. Understanding the energy requirements for each configuration will allow optimization of pacing output.
Methods And Results:
Pacing impedance, LV voltage threshold at 1.5 ms (rheobase) and 0.4 ms and chronaxie were obtained in 6 LV configurations in 49 consecutive patients undergoing CRT implantation or replacement. Strength-duration curves were derived using the Lapicque formula. Pacing impedances and voltage thresholds at 1.5 and 0.4 ms, calculated minimum threshold energy at chronaxie, current drain, energy thresholds at 0.4 ms, and strength-duration curves were statistically different between LV configurations (P<0.05). The lowest threshold energy requirements were found in Tip→right ventricular (RV) coil and Tip→Can configuration. Energy strength-duration curves involving the ring as the cathode (Ring→RV, Ring→Can, and Ring→Tip) had the highest LV thresholds. The pacing configuration with the lowest energy threshold correlated 89% of the time with the lowest voltage threshold at 0.4 ms. The probability to reach LV thresholds <1.5 V at 0.4 ms was increased from 51% with 2 LV configurations to 67% with 6 LV configurations.
Conclusions:
Pacing impedance, LV thresholds, minimum threshold energy at chronaxie, current drain, voltage, and energy strength-duration curves were statistically different between LV pacing configurations. LV pacing configuration with the lowest voltage threshold does not always reflects the lowest energy threshold, particularly in the presence of a low impedance configuration. The availability of 6 LV configurations increases the probability of optimizing LV pacing output.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Cardiomyopathy II: Dilated Cardiomyopathy
Electrophysiology of Normal Cardiac Rhythm
Heart Failure Drugs: Inotropic Agents