Comparison of tools and techniques for implanting pacemaker leads on the ventricular mid-septum
Haran Burri1, Giulia Domenichini, Henri Sunthorn
1Cardiology service, University Hospital of Geneva, Geneva, 4 rue Gabrielle-Perret-Gentil, 1211 Geneva 14, Switzerland. haran.burri@hcuge.ch
Introduction:
Many physicians target the interventricular septum for pacemaker implantation, but the lead may inadvertently end up in an anterior position.
Aims:
We sought to compare two stylet shapes to achieve mid-septal lead placement, as well as the utility of a novel right anterior oblique (RAO) fluoroscopic landmark.
Methods And Results:
Patients undergoing pacemaker implantation were enrolled into four consecutive groups according to stylet shape: a standard curve [two-dimensional (2D) stylet] or with an additional distal posterior curve [three-dimensional (3D) stylet], and whether RAO fluoroscopy was used. Left oblique anterior (LAO) and postero-anterior (PA) fluoroscopic views were used in all cases. After implantation, validation of right ventricular lead position (septal vs. anterior) was performed by echocardiography. A total of 113 patients were included, of whom lead position could be validated in 106 patients. Septal position was achieved in only 10 of 22 (45%) patients in the 2D stylet group and in 17 of 23 (74%) patients in the 3D stylet group (P = 0.07) when only PA and LAO fluoroscopy were used. Results were significantly improved by additional use of RAO fluoroscopy, with successful septal placement in 25 of 28 (89%) patients in the 2D stylet + RAO group (P = 0.001) and 32 of 33 (97%) patients in the 3D stylet + RAO group (P = 0.015).
Conclusions:
A septal lead position was obtained in only about half of the patients when a 2D stylet was used with only LAO and PA fluoroscopic views. A 3D stylet was useful to attain the target position, and additional RAO fluoroscopy significantly improved success rate with both stylet shapes.