Related Experiment Video
Updated: May 26, 2026

An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
Relationship of cortical thickness of the proximal humerus and pullout strength of a locked plate and screw construct
Maegen J Wallace1, Gary Bledsoe, Berton R Moed
1Department of Orthopaedic Surgery, St Louis University, St Louis, MO 63110, USA.
Background:
No study to date has evaluated cortical thickness as it relates to locking plate failure or screw pullout in the proximal humerus. The purpose of this study is to determine the relationship between proximal humerus cortical thickness and locked plate hardware failure in a cadaveric proximal humerus fracture model.
Methods:
Twelve humerus specimens were placed into two groups based on the proximal humerus cortical thickness on an anteroposterior radiograph: less than 4 mm and greater than 4 mm. The specimens were plated with a six-hole proximal humerus locking plate and a 15-mm resection osteotomy at the surgical neck was performed. The specimens were tested in a materials testing machine at a displacement of 5 mm/min to failure.
Results:
Load at failure, stiffness, maximum load, failure, and fracture gap closure were all statistically similar (P > 0.05) between the groups.
Conclusion:
Our biomechanical study used modern locked plate-screw construct fixation of a simulated two-part proximal humerus fracture. The mechanical strength was unaffected based on a threshold combined proximal humerus cortical thickness of 4 mm.
Related Concept Videos
Bones of the Upper Limb: Humerus
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Plastic Deformation in Circular Shafts
Bones of the Upper Limb: Ulna
Screw: Problem Solving
To evaluate the...
Circular Shafts - Elastoplastic Materials
As torque on the...